Funding for this project was provided by the American Academy of Orthotists and Prosthetists through a grant from the US Department of Education (grant number H235K080004). However, this does not necessarily represent the policy of the Department of Education, and you should not assume endorsement by the Federal Government. For more information about the Academy please visit our website at www.oandp.org.



You can help expand the
O&P Virtual Library with a
tax-deductible contribution.

View as PDF

with original layout

Valve Housing for Use in Foaming and Finishing Above-Knee Suction Sockets

Harry N. Hughes, C.P. *
Gene Helmuth *

When foaming and finishing an. above-knee suction socket much time is devoted to achieving an acceptable, finished socket. One of the problems has been to provide a neat cosmetic appearance in the valve area. Another problem has been to provide a properly tapered radius to the valve area so that the valve is easily accessible to the patient. A method used by the staff at the Navy Prosthetics Research Laboratory has proven to be very satisfactory, and is described here.

A plastic valve housing has been designed and is made as a standard production item in two sizes. The larger size has a proximal diameter of 3 cm and distal diameter of 9 cm. It is 7 cm long with a 15-degree offset angle at the proximal end. The smaller housing measures 3 cm in diameter proximally, 5.5 cm distally, and is 3 cm long. The molds used in the manufacture of the plastic valve housing, along with the valve housings, are shown in Fig. 1.

After the socket is laminated, approximately .5 cm of the laminate is removed around the face of the valve to allow placement of the plastic housing, which has been designed to fit snugly over the valve (Fig. 2 ). At this time, the prosthetist visualizes the contour required for finishing, selects the appropriate size housing (depending upon the anticipated depth of the foam), and shapes the housing on the sander (Fig. 3 ). After shaping, the housing is filled with plaster-of-Paris or clay and plastic foam is introduced between the socket and the knee. After the foaming operation the thigh section is given its final shape (Fig. 4 ) and an outer laminate is applied. When the lamination has cured, the socket is trimmed, the filler is removed, and the thigh section is attached to the shin and foot. A view of the final result is shown in Fig. 5 .

This procedure eliminates the need for removing foam from around the valve and the filling in of this area with Cabosil, thus saving time while giving a much neater finish.

The Navy Prosthetics Research Laboratory will be happy to provide samples of these plastic valve housings upon request.